• Skip to main content
  • Skip to footer
CAEN logo light
  • About
    • Company Profile
    • Our policy
    • Sales network
    • Innovative Projects
    • Careers
    • How to reach us
  • Products
  • Resources
    • Documentation
    • Download
    • FAQ
    • Glossary
    • Certifications
    • Legal and Compliance
  • Support
  • Contact
All productsSpecial OffersForm FactorEducationalPower SupplyModular Pulse ProcessingDigitizer FamiliesFERS FamiliesDigital SpectroscopyCAEN SyS productsFirmware & SoftwarePowered CratesAccessoriesBrands
Background image

R8032 8-16 Channel 500 V/10 mA 19'' Power Supply Module (USB/Ethernet/Touchscreen)

Request a Quote Add to WishlistView Wishlist

Home Power SupplyStand Alone Power Supply8/16 Channel up to 6 kV Family with USB, Ethernet & Touchscreen R8032

Photo of R8032
Photo of R8032
DatasheetDownloadCompareTech. Spec.

Highlights

  • 8 or 16 independent channels in a 19″ rack unit (110/220V AC Powered)

  • 500 V / 10 mA output range

  • Channels with either positive or negative polarity

  • SHV coaxial output connectors

  • Low Ripple

  • Under/over-voltage alert, overcurrent and max. voltage protection

  • Interlock logic for board enable and Individual channel kill

  • 50 nA Current monitor resolution (with x10 Imon-Zoom: 5 nA)

  • 2.8″ color touch screen display

  • Local and Remote control (USB2.0/Ethernet)

  • Software Tools for easy channel management

Overview

The CAEN Mod. R8032 provides 8 or 16 independent 500 V / 10 mA channels in a 19″ rack unit (110/220 V AC powered). Each channel delivers up to 500 V and 10 mA. The unit is available with positive, negative or “mixed” channels (4-8 positive and 4-8 negative).

The Mod. R8032 is housed in a 19″ 2U rack unit and, in both the 8- and 16-channel versions, delivers its outputs through SHV connectors; for further technical information, consult our connectors reference page. Ramp-up and ramp-down rates can be set independently for each channel from 1 to 100 V/s in 1 V/s steps, and the module offers 50 nA Iset resolution.

The module can be controlled locally, through a 2.8″ color touchscreen LCD display with a completely redesigned user interface, or remotely via USB 2.0 or Ethernet. A complete set of software tools is available to control the unit, and users can freely download low-level libraries, the LabVIEW driver and graphical application software.

The module also includes dedicated safety features that allow it to operate as a current generator. An overvoltage and undervoltage warning is issued when the output voltage differs from the programmed value by more than 2% of the set value (minimum 10 V). When a channel attempts to exceed its programmed current limit, it is flagged as being in “overcurrent” and enters TRIP status: the output voltage is adjusted to keep the current below the programmed limit for a programmable TRIP time, after which the channel is switched off. If TRIP is set to “constant current mode”, the channel behaves as a current generator. Hardware VMAX and IMAX limits can be set via front panel potentiometers to the same common value for all channels of the board, and both values can be read out via software. Finally, a common safety board interlock logic enables or disables the channels, while individual input signals provide a kill function for each channel.

Supported Software

  • CAEN HV Wrapper Library — Library for CAEN Power Supply Control
  • GECO2020 — GEneral COntrol Software for CAEN HV Power Supplies
  • LabVIEW Driver (PSM – Power Supply Modules) — LabVIEW Instrument Driver for Power Supply Modules

Accessories

A148x
A148x
Inhibit - Kill Signal BNC Adapter for HV Power Supply Modules
HV CABLES
HV CABLES
High Voltage Cable Assemblies

Ordering Options

Code Description
WR8032DXAAAA

R8032DN – 16CH Rack-mount Programmable HV PowerSupply(-500V 10mA) – SHV conn. – Common Gnd

RoHS
WR8032DXMAAA

R8032DM – 16CH Rack-mount Programmble HV P.S.(8ch-500V 10mA,8ch +500V 10mA) – SHV conn. – C. Gnd

RoHS
WR8032DXPAAA

R8032DP – 16CH Rack-mount Programmable HV PowerSupply(+500V 10mA) – SHV conn. – Common Gnd

RoHS
WR8032XAAAAA

R8032N – 8CH Rack-mount Programmable HV PowerSupply(-500V 10mA) – SHV conn. – Common Gnd

RoHS
WR8032XMAAAA

R8032M – 8CH Rack-mount Programmable HV P.S.(4ch -500V 10mA,4ch +500V 10mA) – SHV conn. – Common Gnd

RoHS
WR8032XPAAAA

R8032P – 8CH Rack-mount Programmable HV PowerSupply(+500V 10mA) – SHV conn. – Common Gnd

RoHS

You also may be interested in…

DT8032
DT8032
8 Channel 500 V/ 10 mA Desktop Power Supply Module (USB/Ethernet/Touchscreen)
N8032
N8032
8 Channel 500 V / 10 mA NIM Power Supply Module (USB/Ethernet/Touchscreen)
Download Compare Tech. Spec.

Add to Wishlist
This product is already in your Wishlist
Browse the Wishlist
Request a Quote

Request a Quote

    Submit

    Technical Specifications

    Close

    Packaging

    19” rack (h: 2U; d: 360mm). 110/220V AC Powered

    Output Channels

    • 8-16 channels, SHV connector

    • Positive, Negative or Mixed (4-8 positive and 4-8 negative) Polarity; common ground

    Output Voltage

    0÷500 V

    Max. Output Current

    10 mA

    Max. Ch. Output Power

    5 W

    Vset Resolution

    2mV

    Vmon Resolution

    2mV

    Iset Resolution

    50 nA

    Imon Resolution

    50 nA (high range) / 5nA (low range)

    Vmax

    0 ÷ 500 V

    Vmax resolution

    1 V

    Ramp Up/Down

    1÷100 Volt/s, 1 Volt/s step

    Trip

    Max. time an “overcurrent” can last (seconds). A channel in “overcurrent” works as a current generator; output voltage varies in order to keep the output current lower than the programmed value. “Overcurrent” lasting more than set value (1 to 9999) causes the channel to “trip”. Output voltage will drop to zero either at the Ramp-down rate or at the fastest available rate, depending on Power Down setting; in both cases the channel is put in the off state. If trip= INFINITE, “overcurrent” lasts indefinitely. TRIP range: 0 ÷ 999.9s; 1000 s = Infinite. Step = 0.1 s

    Voltage Ripple

    • 10 – 1000Hz: <4 mVpp typ; <6 mVpp max

    • 1kHz – 20MHz: <2 mVpp typ; <4 mVpp max

    Vmon vs. Vout Accuracy

    ± 0.02% ± 100 mV

    Vset vs. Vout Accuracy

    ± 0.02% ± 100 mV

    Imon vs. Iout Accuracy

    ± 0.5% ± 50 µA

    Iset vs. Iout Accuracy

    ± 0.5% ± 50 µA

    Humidity range

    0 ÷ 80% non condensing

    Operating Temperature

    0 ÷ 45°C

    Storage temperature

    -10 ÷ 70°C

    Vout / Temperature coefficient

    ±50 ppm/°C typ

    Long Term stability (1 week after 1h warmup)

    ±0.04% / full scale

    Local Control

    LCD touchscreen

    Remote Control

    USB & Ethernet

    Footer

    CAEN - Tools for Discovery Logo (RGB)

    CAEN S.p.A.

    PI 00864500467 | REA: LU 102690

    C.I.V.: 500.500€ | CDU A47Ø7H7

    CAEN Group logo (B&W)
    IMQ Certification
    Cyber Essentials Plus

    About Us

    • Company Profile
    • Our policy
    • Sales network
    • Innovative Projects
    • Careers
    • How to reach us

    Resources

    • Documentation
    • Accessibility Statement
    • Legal and Compliance
    • Certifications

    Applications

    • High Energy Physics
    • Astrophysics
    • Neutrino Physics
    • Dark Matter Investigation
    • Nuclear Physics
    • Material Science
    • Medical Applications
    • Homeland Security

    Support

    • Getting Started with MyCAEN Portal
    • Login / Register
    • Wishlist

    News

    • News

    Contact Us

    • Contact
    • Privacy Policy
    • Cookie Policy
    • Whistleblowing

    Copyright © 2026 CAEN S.p.A. All rights reserved. Website by Addiction

    ✕