R1419ET

8/4 Ch Reversible 500 V/200 µA 19' HV Power Supply Module (USB/Ethernet/T.screen)

Datasheet

Home Power SupplyStand Alone Power SupplyUp to 15 kV Reversible Polarity - Rack R1419ET

Photo of R1419ET
Photo of R1419ET
Photo of R1419ET
  • 4 or 8 independent channels in a 19″ rack unit (110/220V AC Powered)

  • 500 V / 200µA output range

  • Channels with individually selectable positive or negative polarity

  • SHV coaxial output connectors

  • Common floating return

  • Low Ripple

  • Under/over-voltage alert, overcurrent and max. voltage protection

  • Interlock logic for board enable and Individual channel kill

  • 5 nA Current monitor resolution (with x10 Imon-Zoom: 500 pA)

  • 2.8″ color touch screen display

  • Local and Remote control (USB2.0/Ethernet)

  • Daisy-chain capability (4 channel only)

  • Software Tools for easy channel management

The CAEN Mod. R1419ET provides 8 or 4 (depending on version) independent High Voltage channels in a 19″ rack unit (110/220V AC Powered). Each channel can provide a ±500V max voltage and a 200 µA max current. The output polarity is independently selectable for each channel.
Channels have common floating return (common return insulated from the crate ground); HV outputs are delivered through SHV connectors. The HV output RAMP-UP and RAMP-DOWN rates may be selected independently for each channel in the range 1÷50 V/s in 1 V/s steps. The module features 5 nA Iset/Imon resolution. Zoom (x 10) for Imon increases resolution to 500 pA.
Module control can take place either locally thanks to a 2.8″ Touchscreen Graphic color LCD display with a completely redesigned user interface or remotely, via USB 2.0 or Ethernet.

SHV connector

19” 2U,  4/8 channels for Mod. R1419ET

Consult our connectors reference page for technical information.

A complete set of free software Tools is available to control this unit: GECO2020 with user friendly GUI and CAEN HV Wrapper library for custom SW development. EPICS and LabVIEW also supported.

Safety features allows the module to perform as a current generator and includes:

Interlock

Common Interlock logic for channels enable/disable and individual inputs signal for channel Kill function.

Overvoltage and Undervoltage

warning when the output voltage differs from the programmed value by more than 2% of set value (minimum 10V).

Overcurrent detection

if a channel tries to draw a current larger than its programmed limit, it enters TRIP status, keeping the maximum allowed value for a programmable time (TRIP), before being switched off.

VMAX

Programmable VMAX protection limit.

Software

CAEN HV Wrapper Library

Library for CAEN Power Supply Control

GECO2020

GEneral COntrol Software for CAEN HV Power Supplies

EPICS IOC (PSM Power Supply Modules)

EPICS IOC for Power Supply Modules

LabVIEW Driver (PSM – Power Supply Modules)

LabVIEW Instrument Driver for Power Supply Modules

Accessories

A148x

A148x

Inhibit - Kill Signal BNC Adapter for HV Power Supply Modules
HV CABLES

HV CABLES

High Voltage Cable Assemblies

Ordering Options

Code Description
WR1419ETDXAA

R1419ETD – 8 Channel 500V/200uA 19″ HV Power Supply Module with Ethernet & 2.8″ Touchscreen

RoHS
WR1419ETXAAA

R1419ET – 4 Channel 500V/200uA 19″ HV Power Supply Module with Ethernet & 2.8″ Touchscreen

RoHS
Download Compare Tech. Spec.

Add to Wishlist
This product is already in your Wishlist
Browse the Wishlist

Request a Quote

Request a Quote

    Submit I’ve read and accept the privacy policy *

    Compare

    Close
    Image
    Name
    field_1
    field_2
    field_3
    field_4
    field_5
    field_6
    field_7
    field_10
    field_8
    field_9
    R1470ET

    R1470ET

    4-8

    ±8000 V

    3 mA

    200 mV

    Iset: 50 nA <br>Imon: 50 / 5 nA

    500 V/s

    <5-10 mVpp

    SHV

    9 W (< 3 kV) 8 W (> 3 kV)

    Remote control via USB or Ethernet

    R1471ET

    R1471ET

    4-8

    ±5500 V

    300 µA

    100 mV

    Iset: 5 nA <br>Imon: 5 / 0.5 nA

    500 V/s

    <3 mVpp

    SHV

    1.65 W

    Remote control via USB or Ethernet

    R1419ET

    R1419ET

    4-8

    ±500 V

    200 µA

    10 mV

    Iset: 5 nA <br>Imon: 5 / 0.5 nA

    100 V/s

    <5 mVpp

    SHV

    0.1 W

    Remote control via USB or Ethernet

    R1570ET

    R1570ET

    2/4

    ±15000 V

    1 mA

    400 mV

    Iset: 20 nA <br>Imon: 20 / 2 nA

    500 V/s

    <4-20 mVpp

    LEMO HV

    10 W

    Remote control via USB or Ethernet

    R1471HET

    R1471HET

    4-8

    ±5500 V

    20 µA

    100 mV

    Iset: 1 nA <br>Imon: 1 / 0.05 nA

    500 V/s

    <2 mVpp

    SHV

    0.11 W

    Remote control via USB or Ethernet

    R1472ETS

    R1472ETS

    1

    ± 3000 V

    2 mA

    100 mV

    Iset: 50 nA, Imon: 5 nA

    1÷500 Volt/s, 1 Volt/s step

    Typ: < 5mVpp; Max: <15mVpp

    SHV

    6 W

    Remote control via USB, Ethernet, T. Screen

    Technical Specifications

    Close

    Packaging

    19” rack (h: 2U; d: 360mm). Weight: ~9kg (2-4 ch), 10.5kg (8 ch);

    Output channels

    4-8 channels, Common Floating Return, SHV connector
    Positive or Negative Polarity (requires internal setting)

    Output ranges

    500 V / 200 µA (IMonRange = High)
    500 V / 20 µA (IMonRange = Low) – Imon Zoom Active

    Max. Ch. Output Power

    0.1 W

    Vset Resolution

    10 mV

    Vmon Resolution

    10 mV

    Iset Resolution

    5 nA

    Imon Resolution

    5 nA (IMonRange = High)
    500 pA (IMonRange = Low) – Imon Zoom Active

    Vmax

    0 ÷ 510 V Absolute maximum HV level that the channel is allowed to reach, independently from the preset value Vset.
    Output voltage cannot exceed the preset value Vmax. The accuracy is 1 % ± 5 V

    Vmax resolution

    ± 0.1 V

    Alarm output

    Open collector, 100 mA maximum sink current

    Interlock input

    LOW: < 1 V; current~5mA; HIGH: 4÷6 V

    Ramp Up/Down

    1÷50 Volt/s, 1 Volt/s step

    Trip

    • Max. time an “overcurrent” can last (seconds). A channel in “overcurrent” works as a current generator; output voltage varies in order to keep the output current lower than the programmed value.
    • “Overcurrent” lasting more than set value (1 to 9999) causes the channel to “trip”.
    • Output voltage will drop to zero either at the Ramp-down rate or at the fastest available rate, depending on Power Down setting; in both cases the channel is put in the OFF state.
    • Trip range: 0 ÷ 999.9 s; 1000 s = INFINITE. Step = 0.1 s (If trip= INFINITE, “overcurrent” lasts indefinitely)

    Voltage Ripple

    • 20 ÷ 1000 Hz: Typical: 5 mVpp / Maximum: 10 mVpp
    • 1 ÷ 20000 kHz: Typical: 3 mVpp / Maximum: 5 mVpp

    Vmon vs. Vout Accuracy

    ±0.02% of read value ±0.2 V

    Vset vs. Vout Accuracy

    ±0.02% of read value ±0.2 V

    Imon vs. Iout Accuracy

    ±2% of read value ±20 nA (IMonRange = High)
    ±2% of read value ±2 nA (IMonRange = Low) – Imon Zoom Active

    Iset vs. Iout Accuracy

    ±2% of read value ±30 nA (IMonRange = High)
    ±2% of read value ±3 nA (IMonRange = Low) – Imon Zoom Active

    Ventilation Fan

    60×60 24V; 62 dBA maximum noise level

    Humidity range

    0 ÷ 80%

    Storage temperature

    -10 ÷ 70°C

    Operating temperature

    0 ÷ 45°C

    Vout / Temperature coefficient

    max. 50 ppm / °C

    Imon / Temperature coefficient

    max 100 ppm / °C; max 300 ppm / °C with Imon zoom

    Long term stability Vout vs. Vset

    ± 0.02% (after one week @ constant temperature)

    Footer

    CAEN S.p.A.

    PI 00864500467 | REA: LU 102690

    C.I.V.: 500.500€ | CDU A47Ø7H7

    IMQ Certification
    Cyber Essentials Plus

    About Us

    • Company Profile
    • Our policy
    • Sales network
    • Innovative Projects
    • Careers
    • How to reach us

    Resources

    • Documentation
    • Accessibility Statement

    Applications

    • High Energy Physics
    • Astrophysics
    • Neutrino Physics
    • Dark Matter Investigation
    • Nuclear Physics
    • Material Science
    • Medical Applications
    • Homeland Security

    Support

    • Getting Started with MyCAEN Portal
    • CAEN LHC Commitment

    News

    • News

    Contact Us

    • Contact

    Login

        
    • Privacy Policy
    • Cookie Policy

    Copyright © 2026 CAEN S.p.A. All rights reserved. Website by Addiction

    ✕